The V-1100 is an automated hand-held evidential Breath Alcohol Analyzer.
The V-1100 Alcohol Tester follows a program driven test procedure.
Each function is displayed which provides the operator with step-by-step
instructions.
The instrument utilizes fuel cell to quantify alcohol in the provided
Breath sample. Fuel cell is mounted in a valve assembly that is
designed to draw a fixed volume sample across the surface of the
fuel cell. When a subject’s breath sample is drawn across the top
surface of the cell, any alcohol that is broken down a given number
of electrons are freed during the process. The number of electrons
(the resulting electrical signal) a digital BrAc (Breath Alcohol
Concentration) is calculated and reported. |
|
|
SPECIFICATION |
|
| Sensor |
: |
Fuel
cell |
| Machine
weight (without battery) |
: |
240gm |
| Battery
|
: |
Four
AA batteries (alkaline) |
| External
power supply |
: |
12
VDC (Car Battery) |
| Battery
use Life |
: |
Not
less than 300 times (high quality AA type alkaline battery) |
| Response
time |
: |
<25
seconds |
| Benchmark
work condition |
: |
|
| Temperature |
: |
25
± 5° C |
| Relative humidity |
: |
No higher than 90% |
| Air pressure |
: |
86 Kpa to 106 Kpa |
| Operating
Temperature |
: |
-20°-70°C |
| Storage
temperature |
: |
-5°-50°C |
| Storage
relative humidity |
: |
No
higher than 90% |
| Testing
bias |
: |
±
10% (Range: 0.400mg/1-1.000mg/1)
± 0.040mg/1(Range:0.200mg/1-2-0.400mg/1)
± 0.025 mg/1(Range:0.400mg/1-1.000mg/1) |
| Testing
range |
: |
0.00mg
/1-1.00mg/1 (0-400mg/100ml) |
| Language |
: |
English |
| Display |
: |
128X32Matrix LCD screen |
| Printer
output |
: |
RS232 |
| Printer
(Optional) |
: |
Dot matrix
24 col. |
| Complete
Package |
: |
1. Detector
Instrument V 1100
2. Printer (with built in rechargeable battery)
3. Charger
4. Printer Cable
5. Mouth Piece
6. Carring case
7. Instruction manual
|